Decay-Delay Technology for Management of Defibrillator T-wave Oversensing

AUTHORS

Amir Aslani 1 , * , M Moradi 2 , J Kheirkhah 2 , M Kafi 2

AUTHORS INFORMATION

1 Cardiovascular Research Center, Shiraz University of Medical Sciences, Shiraz, Iran

2 Shaheed Rajaei Cardiovascular Medical Centre, Tehran University of Medical sciences, Tehran, Iran

How to Cite: Aslani A, Moradi M , Kheirkhah J , Kafi M . Decay-Delay Technology for Management of Defibrillator T-wave Oversensing, Int Cardio Res J. 2010 ; 4(4):e61870.

ARTICLE INFORMATION

International Cardiovascular Research Journal: 4 (4); e61870
Published Online: December 01, 2010
Article Type: Case Report
Received: September 20, 2017
Accepted: December 01, 2010

Crossmark

CHEKING

READ FULL TEXT
Abstract

Appropriate sensing is indispensable for appropriate functioning of implantable cardioverter-defibrillator (ICD). Oversensing is a major clinical dilemma in patients with ICD leading to inappropriate ventricular arrhythmia detection and therapy. T-wave oversensing is an important clinical problem leading to QRS double counting with inappropriate therapy, in particular. We present a patient with dilated cardiomyopathy and intermittent T wave oversensing by an ICD which could be resolved by device reprogramming.